Home / all / In-vitro Diagnostic reagents / Diagnostic Apparatuses / Chemistry Analyzer /

CHINA BKE Series Auto Electrolyte Analyzer Urine Test Strips Analyzer For Medical And Laboratory

Categories

CHINA BKE Series Auto Electrolyte Analyzer Urine Test Strips Analyzer For Medical And Laboratory

CHINA BKE Series Auto Electrolyte Analyzer Urine Test Strips Analyzer For Medical And Laboratory
CHINA BKE Series Auto Electrolyte Analyzer Urine Test Strips Analyzer For Medical And LaboratoryCHINA BKE Series Auto Electrolyte Analyzer Urine Test Strips Analyzer For Medical And LaboratoryCHINA BKE Series Auto Electrolyte Analyzer Urine Test Strips Analyzer For Medical And LaboratoryCHINA BKE Series Auto Electrolyte Analyzer Urine Test Strips Analyzer For Medical And LaboratoryCHINA BKE Series Auto Electrolyte Analyzer Urine Test Strips Analyzer For Medical And Laboratory
CategoriesChemistry Analyzer
ModelBKE
TypeBlood & Gas Analysis System
Instrument classificationClass II
ModelBKE Series
Analysis Methodlon selective electrode (ISE)
Storageup to 10,000 test results
FOB portTianjin xingang
Terms of PaymentL/C, D/A, D/P, Western Union, MoneyGram, T/T, Paypal
Update Time2022/5/18
Detail Information
CHINA BKE Series Auto Electrolyte Analyzer Urine Test Strips Analyzer For Medical And Laboratory
Quick Details
Place of Origin
China
Model Number
BKE
Instrument classification
Class II
Sample
Serum, plasma, whole blood, cerebrospinal fluid and dilute urine
Measuring Speed
≤25s
Analysis Method
lon selective electrode (ISE)
Sample Volume
60-300ul (3 parameters to 11 parameters)
Sample Position
35 Positions (including 1 QC)
Storage
up to 10,000 test results
Printer
Built-inthermal printer
Interface
RS232 port
Power Supply
AC220±10% 50/60Hz; 110±10%,60Hz; 120W
BKE Series Auto Electrolyte Analyzer

The instrument is intended use in hospitals to measure K+、Na+、Cl-、Ca2+、Li ions and TCO2 as well as pH values, AG in samples of whole blood, serum, plasma, urine solution(diluted).
Product Description
Features
 1.  Automatic electric potential tracking and correcting software . 
 2.  Automatically detect and filter tiny bubbles. 
 3.  Real-time diagnostic of system working status. 
 4.  Waste liquid automatic detection and alarming.
 5.  Automatic calibration and two-point correction.
 6.  Wave theory flushing method and direct flushing pipe method.
 7.  High accurate and long life electrode and TCO2 sensor.
 8.  Sleep mode to reduce reagent consumption.
 9.  Rapid test speed of 80 tests per hour.
 10. Option: Sample tray.
Specification
Model
Measuring Range
Resolution
Measuring Precision
K+
0.5~20.0mmol/L
0.01mmol/L
≤1.0%
Na+
15~200mmol/L
0.1mmol/L
≤1.0%
Cl-
15~200mmol/L
0.1mmol/L
≤1.0%
Ca2+
0.1~6.0mmol/L
0.01mmol/L
≤1.0%
Li+
0.1~5.0mmol/L
0.01mmol/L
≤2.0%
PH
4~9pH
0.01pH
≤5.0%
TCO2
2.0~70.0mmol/L
0.1mmol/L
≤3.0%
Model
Test Items
BKE-A
K, Na, Cl
BKE-B
K, Na, Cl, TCO2, AG
BKE-C
K, Na, Cl, iCa, nCa, TCa, pH
BKE-D
K, Na, Cl, iCa, nCa, TCa, pH, TCO2, AG
BKE-F

K, Na, Cl, Li
BKE-H
K, Na, Cl, iCa, nCa, TCa, pH, Li
BKE-I
K, Na, Cl, iCa, nCa, TCa, pH, Li, TCO2, AG
BKE-J
K, Na, Cl, Mg
BKE-K
K, Na, Cl, iCa, nCa, TCa, pH, Mg
BKE-L
K, Na, Cl, iCa, nCa, TCa, pH, Li, Mg, TCO2, AG
Model
BKE Series
Sample
Serum, plasma, whole blood, cerebrospinal fluid and dilute urine
Measuring Speed
≤25s
Analysis Method
lon selective electrode (ISE)
Sample Volume
60-300ul (3 parameters to 11 parameters)
Sample Position
35 Positions (including 1 QC)
Storage
up to 10,000 test results
Printer
Built-inthermal printer
Interface
RS232 port
Working Conditions
Power Supply
AC220±10% 50/60Hz; 110±10%,60Hz; 120W
Temperature
5-40℃
Relative Humidity
≤ 80 %
Atmospheric Pressure
(86~106) kPa
Package Size(W*D*H)
490*390*700mm
Gross Weight
27kg
Please send your message to us
*Email
Name
Tel
*Title
*Content
Address
Categories
KEEP IN TOUCH
Please send your message to us
*Email
Name
Tel
*Title
*Content
Address
Contact us
subscription